Binder KBF 130 constant climate chamber
€16.148,00 excl. VAT
Benefits:
- A reliable option for critical ICH stability testing thanks to maximum climate homogeneity, even when the unit is fully loaded
- Reliable: failsafe operation over the long term thanks to proven compressor technology and interior made of stainless steel
- Smart: signal triangle with light-up status display and ergonomic control terminal with touchscreen program controller
- Economical: up to 40% lower energy consumption than the previous model thanks to modern inverter cooling technology
Important features:
- Temperature range: 0°C to +70°C
- Humidity range: 10% to 80% RH
- Proven APT.line™ preheating chamber technology
- Up to 40% lower energy consumption compared to the previous model
- Energy-efficient cooling with modern inverter cooling unit
- Cooling with climate-neutral R-600a refrigerant
- Door heating to prevent condensation
- Adjustable fan speed
- Humidity regulation with capacitive humidity sensor and responsive steam humidification
- Robust housing design with HIT sandwich technology
- Stainless-steel inner chamber
- Inner doors made of metal oxide-coated safety glass
- Two racks made of stainless steel
- Large access port with silicone plug
- Tabletop units (up to model 130) are stackable with stacking adapter
- Floor-standing units (from model 260) featuring robust twin casters, with brakes at the front
- Intuitive touchscreen controller with time-segment and real-time programming
- Ergonomically adjustable control terminal (from model 470)
- Signal triangle with light-up status display (from model 470)
- Internal data logger, measured values can be read out in open format via USB
- Class 3.3 independent temperature safety device (DIN 12880) with visual and acoustic alarms
- Unit self-test for comprehensive status analysis
- Data interfaces: Ethernet, USB
- BINDER Multi Management Software APT-COM™ basic edition
- Optional light modules for ICH photostability testing (for models 260, 470, 720)
- Optional LED light modules for plant growth (for models from 260)
Description

Binder KBF 130 Constant Climate Chamber – Compact Precision for Stability Testing
The Binder KBF 130 constant climate chamber is a compact, high-performance solution for ICH-compliant stability testing in pharmaceutical and research laboratories. With a 127-liter stainless steel interior, it offers precise temperature control from 0 °C to +70 °C and humidity regulation from 10% to 80% RH.
This model features Binder’s advanced APT.line™ preheating chamber technology and energy-efficient inverter compressor cooling using climate-friendly refrigerant R-600a. The result is uniform climate conditions throughout the chamber, even under full load, and up to 40% lower energy consumption compared to previous models.
The KBF 130 is ideal for long-term testing, shelf-life studies, and quality assurance applications. Its modern touchscreen controller allows for intuitive programming, while integrated data logging and Ethernet/USB interfaces support GLP-compliant documentation. Safety is ensured with a class 3.1 independent temperature safety device and door heating to prevent condensation.
Key Features:
-
Temperature range: 0 °C to +70 °C
-
Humidity range: 10% to 80% RH
-
Volume: 127 liters
-
APT.line™ preheating for uniform conditions
-
Inverter compressor with R-600a refrigerant
-
Touchscreen controller with real-time monitoring
-
Internal data logger with USB/Ethernet
-
Class 3.1 safety device and door heating
-
Stackable, space-saving design
The Binder KBF 130 climate chamber offers precision, reliability, and energy efficiency in a compact footprint—perfect for controlled, reproducible test environments.
Specifications
Warranty: 2 years
Lead time: 1-2 weeks
Designation | KBF130-230V |
Article Number | 9020-0478 |
Performance Data Temperature | |
Temperature range | 0…70 °C |
Max. heat compensation at 40°C | 270 W |
Performance Data Climate | |
Temperature range with humidity | 10…70 °C |
Humidity range | 10…80 % RH |
Temperature uniformity at 25°C and 60% RH | 0.2 ± K |
Temperature uniformity at 40°C and 75% RH | 0.2 ± K |
Temperature fluctuation at 25°C and 60% RH | 0.1 ± K |
Temperature fluctuation at 40°C and 75% RH | 0.1 ± K |
Recovery time after door was opened for 30 s at 25°C and 60% RH | 3 min |
Recovery time after door was opened for 30 s at 40°C and 75% RH | 3 min |
Measures | |
Interior volume | 127 L |
Net weight of the unit (empty) | 126 kg |
Load per rack | 25 kg |
Permitted load | 50 kg |
Doors | |
Unit doors | 1 |
Inner doors | 1 |
Internal Dimensions | |
Interior width | 656 mm |
Interior height | 500 mm |
Interior depth | 386 mm |
Wall clearance | |
Wall clearance back | 100 mm |
Wall clearance sidewise | 100 mm |
Housing dimensions not incl. fittings and connections | |
Width net | 900 mm |
Height net | 1,005 mm |
Depth net | 710 mm |
Electrical data | |
Housing protection acc. to EN 60529 | IP 20 |
Rated Voltage | 200…240 V |
Power frequency | 50/60 Hz |
Nominal power | 1,3 kW |
Nominal current | 5,4 A |
Unit fuse | 16,0 A |
Phase (Nominal voltage) | 1~ |
Environment-specific data | |
Schalldruckpegel bei 25°C und 60 % r.F. | 47 dB(A) |
Energy consumption at 25°C and 60% RH | 160 Wh/h |
Energy consumption at 40°C and 75% RH | 160 Wh/h |
Energy consumption at 85°C and 85% RH | |
Fixtures | |
Number of shelves (std./max.) | 2/6 |
Full list: see datasheet