Binder WIC 1 walk-in climate chamber

Pricing on request *

Benefits: 

  • Stable and uniform environment
  • Fully complies with Pharma regulations (ICH)Equipment
  • Longevity – corrosion resistant
  • Ease of operation
  • Environmental friendly refrigerant with GWP value 547
  • Maintenance possible with the least possible disruption to the climate conditioning
  • Perfectly meets requirements in Pharma industry
  • 3 years warranty

Important features:

  • Relative humidity accuracy of ± 2.5%
  • Temperature accuracy of ± 1.5°C
  • Duct work for proper distribution of air
  • Highly insulated room panels with U value of only 0,21 W/m2K
  • LED Lighting with motion detector
  • High quality room elements, made in Germany
  • Interior completely stainless steel
  • Exterior service accessibility
  • Room Elements: Galvanized sheet steel, polyester coated, white, similar to RAL 9010, interior surfaces in stainless steel, 4 inch / 100 mm PU foam U value: 0,21W/m2K
  • Floor: Floor overlay is chrome-nickel steel, pattern-rolled 5WL
  • Surface load: 50.000 N/m2

 

Description

label climate chambers made in germany

Binder WIC 1 Walk‑In Chamber – 12 m³ Walk‑In Environmental & Stability Chamber

The Binder WIC 1 walk‑in chamber delivers precise environmental control for ICH‑compliant stability testing, shelf‑life studies, and accelerated aging. With a generous interior volume of 12 m³ (2,400 × 2,300 × 2,270 mm), it maintains temperatures between 10 °C and 50 °C (±1.5 °C) and humidity from 20 % to 90 % RH (±2.5 %) via a high-efficiency ducted air distribution system. The interior is fully stainless steel, while the exterior is constructed from galvanized and polyester-coated panels in RAL 9010, featuring 100 mm polyurethane insulation with a U-value of 0.21 W/m²K.

This walk-in environmental chamber includes LED lighting with motion sensors, a lockable safety door, a 50 mm access port for external sensors, and an externally installed air-conditioning unit for hassle-free maintenance. Designed for stability studies in pharmaceutical, food, cosmetic, and materials testing labs, the WIC 1 supports compliance with ICH Q1A (R2), GMP, and GLP guidelines.

Equipped to handle floor loads up to 50,000 N/m² and using environmentally friendly refrigerant (GWP 547), the WIC 1 offers robust performance, ease of servicing, and long-term reliability. A 3-year warranty is included.

Key Features:

  • Volume: 12 m³ (2,400 × 2,300 × 2,270 mm)

  • Temperature Range: 10 °C to 50 °C (±1.5 °C)

  • Humidity Range: 20 % to 90 % RH (±2.5 %)

  • Construction: Stainless steel interior, insulated RAL 9010 exterior panels

  • Insulation: 100 mm PU panels, U-value 0.21 W/m²K

  • Climate Uniformity: Ducted air system

  • Lighting: LED with motion sensor

  • Included Features: Lockable door, 50 mm access port, optional ramp and shelving

  • Maintenance: External unit servicing avoids interior climate disruption

  • Compliance: ICH Q1A (R2), GLP, GMP

  • Warranty: 3 years

Specifications

Warranty: 3 years

Lead time: 8 weeks

Designation WIC1-230V
Article Number 9800-0011
Performance Data Climate
Temperature range with humidity 10…50 °C
Humidity range 20…90 % RH
Temperature uniformity at 25°C and 60% RH 1.5 ± K
Temperature uniformity at 40°C and 75% RH 1.5 ± K
Temperature fluctuation at 25°C and 60% RH 0.5 ± K
Temperature fluctuation at 40°C and 75% RH 0.5 ± K
Measures
Interior volume 12,530 L
Net weight of the unit (empty) 776 kg
Doors
Unit doors 1
Internal Dimensions
Interior width 2,400 mm
Interior height 2,270 mm
Interior depth 2,300 mm
Housing dimensions not incl. fittings and connections
Width net 2,600 mm
Height net 2,470 mm
Depth net 2,500 mm
Electrical data
Housing protection acc. to EN 60529 IP 20
Rated Voltage 200…230 V
Power frequency 50 Hz
Nominal power 2,7 kW
Nominal current 11,0 A
Unit fuse 16,0 A
Phase (Nominal voltage) 1~
Environment-specific data
Energy consumption at 40°C and 75% RH 1,130 Wh/h

 

Full list: see datasheet